Dansensor CheckPoint 4 O2, O2/CO2 Portable Headspace Gas Analyser
Portable and accurate headspace gas analyser, combining accuracy with convenience.
The Dansensor CheckPoint 4 is robust and portable, incorporating a choice of oxygen sensors, to best suit individual applications. Select either:
- “O” Version – a new accurate optical oxygen sensor, with a full measurement range of 0 – 100% O2. Recommended for the vast majority of applications, except those where alcohol is present
- “EC” Version – a fast, accurate electrochemical oxygen sensor with a full measurement range of 0 – 100% O2 Recommended for applications where alcohol is present in the headspace.
Both sensor types are available in both CheckPoint 4 versions:
Dansensor CheckPoint 4 Premium
- %O2 & %Balance – choose the O2 version
- %O2 & %CO2 and %Balance – choose the O2 /CO2 version
CheckPoint 3 Basic
- %O2 & % Balance – choose the O2 version
- %O2 & %CO2 & % Balance – choose the O2 /CO2 version
Choose how you use the test data
- WiFi enabled for automatic data download & logging – CheckPoint 4 Premium
- Visual Display only and hard copy transcribed by operator – CheckPoint 4 Standard
Small sample volume. Only requires a 5ml gas sample to complete an analysis. Test results are displayed within seven seconds. Ideal for high frequency testing and for testing packs with a small headspace.
Rechargeable batteries. With more environmentally friendly, rechargeable lithium batteries, the unit holds sufficient charge for 2000+ measurements.
3.5” colour touchscreen user interface. “Low gas flow” alarm warns the operator of a blocked water trap, or test needle that needs changing.
Supplied in a carry case, including: 100 septums, 10 needles, 10 water trap filters, power supply.
CheckPoint 3 Premium | CheckPoint 3 Basic |
---|---|
WiFi enabled | Not available |
User set up | Not available |
Gas flow alarm | Gas flow alarm |
Up to 10 individual user IDs | Single user |
100 different product IDs | Single product ID |
Up to 500 test results per product | Single reading |
Dimensions and Weight | 77 x 100 x 175 mm (H x W x D) 0.7Kg |
---|---|
Power Supply | Battery; up to 2000 measurements on fully charged battery at 20°C; recharge time 5 hours max |
Options | Standard consumables kit (1 sampling kit, 1000 septums, 10 needles (0.8 diameter) 10 water trap filters), metal needles |
Connections | Wi-Fi (WPA, WPA2, WPA enterprise) |
Calibration and Service | “O” Version - 12 months. “EC” Version - 6 months |
Working temperature | 0 to +40°C, less than 95% RH non condensing |
Compatibility | CE, RoHS |
What's included in the box? | 100 septums, 10 needles (0.8 diameter), 10 water trap filters, power supply with interchangeable plugs. |
Available Configurations | O2 (Optical sensor) | O2 (Electrochemical sensor) | IR CO2 Sensor (infrared single beam) |
---|---|---|---|
Sample volume (min) | 5ml | 5ml | 5ml |
Sample time (min) | 7 seconds | 7 seconds | 7 seconds |
Measuring range | 0-100% | 0-100% | 0-100% |
Resolution | 0.01% oxygen | 0.01% oxygen | 0.01% carbon dioxide |
Sensor accuracy | ±(0.25% O2 + 2% of readout) | ±0.25% O2 + 2% of readout) | ±2% CO2 in range 0-20% ±3% CO2 in range 20-100% |
Food applications | Most except those containing alcohol | Most including those containing alcohol | All |